SKANAIR® HFC-SH

FEATURES OF SAFETY WEIGHING CABINET SKANAIR® HFC-SH

  • very low noise level during operation, only 52 dB (A)!
  • ceramic working surface
  • increased safety due to two consecutive filters (HEPA H14)
  • the positive pressure plenum ensures that the pre-filter is dynamically sealed and the contaminated air is fed exclusively through the particle filter
  • the increasing pressure drop across the filters is adjusted automatically
  • main filter can be qualified in situ
  • innovative bag-out filter system
  • tested personal protection in accordance with EN 14175-3 for laboratory fume cupboards

» PDF download

EXPLANATION OF AIR GUIDANCE

The uni-directional air flow is guided from the front to the back through HEPA H14 filters. This prevents, that the released particles are leaving the controlled working area.

 

Two consecutive HEPA H14 filter remove particles effectively.

Dimensions in mm W x D x H ( without support frame ) HFC-SH 900 HFC-SH 1200 HFC-S 1300
Outer dimensions W x D x H ( without support frame ) 900 x 740 x 1075 1200 x 740 x 1075 1215 x 880 x 14610
Height with active carbon kit 1235 1235 1815
Inner dimensions / Working area W × D × H 835 x 545 x 600 1135 x 545 x 600 1135 x 655 x 900
Height of front opening at:
Closed front panel
Open front panel
270
470
270
470
545
835